U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TAPERFILL HIP; TAPERFILL HIP STEM, STANDARD, SIZE 11

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TAPERFILL HIP; TAPERFILL HIP STEM, STANDARD, SIZE 11 Back to Search Results
Model Number 425-96-011
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/08/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to patient being in pain and having an infection.
 
Additional Manufacturer Narrative
The reason for this revision surgery was an infection.The previous surgery and the revision detailed in this investigation occurred 8.8 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported components used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports associated with the components that may have contributed to pain or infection.The device and its applicable concomitant devices were verified to have gone through acceptable sterilization processes and were within their respective expiration dates at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There were no findings during this investigation that indicate that the reported device was the source of the pain or had a direct connection with the patient's infection.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERFILL HIP
Common Device Name
TAPERFILL HIP STEM, STANDARD, SIZE 11
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8289853
Report Number1644408-2019-00070
Device Sequence Number11378261
Product Code LPH
UDI-Device Identifier00888912078566
UDI-Public(01)00888912078566
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K130099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/28/2020
Device Model Number425-96-011
Device Catalogue Number425-96-011
Device Lot Number190R1010
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/08/2019
Supplement Date Received by Manufacturer01/31/2019
Initial Report FDA Received Date01/30/2019
Supplement Report FDA Received Date02/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-402, LOT 871B1111; 430-99-056, LOT 607G1065; 932-40-756, LOT 815N1964; 400-03-402, LOT 871B1111; 430-99-056, LOT 607G1065; 932-40-756, LOT 815N1964
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
-
-