Model Number 110-02-035
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/08/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to an infection.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
The reason for this revision surgery was an infection.The previous surgery and the revision detailed in this investigation occurred 1.7 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported components used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports associated with the components that may have contributed to an infection.The device and its applicable concomitant devices were verified to have gone through acceptable sterilization processes and were within their respective expiration dates at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.If information regarding cultures identified in the infection, the severity of the infection or any other relevant information is submitted at a future date, this investigation will be re-evaluated.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's infection.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3D KNEE
Common Device Name
INSERT, 3D EX SZ 8LT 9MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8289996
Report Number1644408-2019-00074
Device Sequence Number277927
Product Code OIY
UDI-Device Identifier00888912108300
UDI-Public(01)00888912108300
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K091956
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/18/2021
Device Model Number110-02-035
Device Catalogue Number391-09-708
Device Lot Number285G1937
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/08/2019
Supplement Date Received by Manufacturer01/31/2019
03/11/2019
Initial Report FDA Received Date01/30/2019
Supplement Report FDA Received Date02/05/2019
03/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
010-55-020, LOT 005A1180; 010-55-020, LOT 005A1180; 010-55-020, LOT 005A1186; 010-55-020, LOT 005A1186; 010-55-025, LOT 006A1260; 010-55-025, LOT 006A1260; 010-55-025, LOT 006A1274; 010-55-025, LOT 006A1274; 110-02-035, LOT 397A1037; 236-01-108, LOT 952A1080; 236-01-108, LOT 952A1080; 336-01-108, LOT 313B1017; 336-01-108, LOT 313B1017; 391-09-708, LOT 285G1937; 010-55-020, LOT 005A1180; 010-55-020, LOT 005A1186; 010-55-025, LOT 006A1260; 010-55-025, LOT 006A1274; 236-01-108, LOT 952A1080; 336-01-108, LOT 313B1017; 391-09-708, LOT 285G1937
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age58 YR
-
-