(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Investigation summary: it was reported performance breakage suture.An empty labeled winding former and a suture in two sections were returned for analysis.During the visual inspection of the sample (two suture pieces), the ends present elongation and were noted cut appears to be by use of a surgical instrument.A functional test was performed in a suture piece and the tensile force were above the minimum requirements.In handling this or any other suture material, care should be taken to avoid damage from handling.Avoid crushing or crimping damage due to application of surgical instruments such as forceps or needle holders.According to the sample condition, suggest an improper handling.Ten unopened samples were returned for analysis.During the visual inspection of the ten samples, no defects were found on the packages.The samples were opened, and the swage and attachment area of needles were as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were observed.A functional test was performed and the tensile force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the representative samples, no breakage suture was found on the sutures and the tested samples met the finished goods requirements.
|