| Model Number |
421614 |
| Medical Device Problem Code |
Misassembled (1398)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheters were loaded into the packaging backwards.The patient advised that when he opened the package he had to touch the entire catheter to get the insertion side.The patient reported that he did not use the product due to the fear of infection.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheters were loaded into the packaging backwards.The patient advised that when he opened the package he had to touch the entire catheter to get the insertion side.The patient reported that he did not use the product due to the fear of infection.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed manufacturing related since the packaging was done by the manufacturer.Visual evaluation of the returned sample noted five clean cath intermittent catheters within the original packaging.The catheters were all noted to be packaged with the tips pointing away from the chevron end.A catheter without the tip in the chevron seal is out of specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿visually inspect the product for any imperfections or surface deterioration prior to use.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.".
|
| |
|