• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-04
Device Problems Difficult to Fold, Unfold or Collapse (1254); Positioning Problem (3009)
Patient Problem No Code Available (3191)
Event Date 01/10/2019
Event Type  Injury  
Event Description
The recipient reportedly experienced a tip fold over of the electrode array.During initial implant surgery, the electrode had advanced slightly off the stylet.The surgeon reloaded the array without issue and inserted the array reportedly without resistance.After an x-ray revealed a tip fold over of the electrode array, the surgeon reopened the incision and attempted to reinsert the array, but the array came out of the cochlea.The recipient's device was explanted and another advanced bionics cochlear device was implanted.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient's activation went well.The device passed the external visual and photographic imaging inspections.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.
 
Manufacturer Narrative
Advanced bionics does not consider this event as reportable.This is considered an operating room return.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
MDR Report Key8290935
MDR Text Key134564207
Report Number3006556115-2019-00022
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016841002
UDI-Public(01)07630016841002(11)180828(17)210831
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/31/2021
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/30/2019
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received02/18/2019
03/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age24 MO
-
-