| Model Number |
PCB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that lens was explanted from patient¿s right eye in secondary procedure because of patient experiencing blurry vision.There was no vitrectomy, no incision enlargement, no sutures required.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation; returned to manufacturer on: 2/5/2019.Device evaluation: the product was returned to the manufacturing site for evaluation.The sample was received in a specimen cup; it was observed under microscope.The lens was returned cut in two pieces since the explanation process.This complaint is related to a patient condition and could not be verified.A product quality deficiency was not identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints related to this production order was conducted in the complaint system.The search revealed that no other complaint has been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|