| Model Number |
8888145044P |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during treatment, blood leak was noted coming from the distal junction of the venous branch.It was stated that there was a break at this place.Treatment was not initiated as there was a risk of air entering the leak and pulmonary embolism.The venous branch was repaired with the repair kit and a dose of antibiotics was given as an intervention.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during treatment, blood leak was noted coming from the distal junction of the venous branch.It was stated that tego adapter was used to tighten the adapters and the cleaning agent used was chlorhexidine 2% alcohol 70 %.It was stated that there was a break at this place.Treatment was not initiated as there was a risk of air entering the leak and pulmonary embolism.The venous branch was repaired with the repair kit and a dose of antibiotics was given as an intervention.It was also stated that it was a luer adapter issue as the small leak came from under the white rubber part.There was no significant blood loss due to theleak.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|