Model Number G35536
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, a patient of unknown age and gender underwent a superficial femoral artery (sfa) procedure in which an advance 35 lp low profile balloon catheter was used.The complaint device was inflated to 11 or 12 atmospheres which was just above nominal pressure and the balloon burst.A ratio of contrast to saline equal to 50/50 was used for inflation.No information was provided regarding type of inflation device.The patient's anatomy was calcified but not angulated.The complaint device was removed intact with no harm to the patient.Another balloon device (manufacturer not specified) was used and the procedure was completed.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Information received (b)(6) 2019 confirms that the device associated with this event did not rupture, but rather experienced difficult advancement through a sheath.This event is no longer reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury , or product malfunction.There is no precedence of this malfunction leading to an adverse event.No additional follow-up reports will be submitted under report reference number 1820334-2019-00238.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Information was received (b)(6) 2019 clarifying the event.It was initially reported that two separate pta devices were used in two separate procedures.However, the cook area representative has since corrected these events.In the first case (reported in 1820334-2019-00239), the pta rupture is accurate.However, the second complaint has been updated to reflect a difficult advancement/insertion of the second pta device into a 5fr cook sheath (this report).The 5fr sheath was not returned, only the second pta device that would not advance.It should be noted, the pta device is compatible with a 6fr sheath.This event is no longer reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury , or product malfunction.There is no precedence of this malfunction leading to an adverse event.No additional follow-up reports will be submitted under report reference number 1820334-2019-00238.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8293626
Report Number1820334-2019-00238
Device Sequence Number1733890
Product Code LIT
UDI-Device Identifier10827002355368
UDI-Public(01)10827002355368(17)211101(10)9278909
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K132020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2021
Device Model NumberG35536
Device Catalogue NumberPTA5-35-135-6-20.0
Device Lot Number9278909
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2019
Initial Date Received by Manufacturer 01/23/2019
Supplement Date Received by Manufacturer04/02/2019
Initial Report FDA Received Date01/31/2019
Supplement Report FDA Received Date05/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COOK RAABE SHEATH 6FR 55CM
-
-