| Model Number |
MMT-523NAP |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump received a replace battery alert without a low battery alert.The customer¿s blood glucose level was unknown.The customer stated the type of battery in use was aaa (b)(6).The customer stated that the battery was not previously used.Customer stated the battery cap was not loose, cracked or damaged.The customer stated that this was the second occurrence of replace battery alert without a low battery warning.The customer also reported repeated failed battery alerts.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to corroded battery tube spring contact and corroded battery tube.Unable to perform all functional testing including the operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime and excessive no delivery test or verify unexpected battery power loss
due to blank display.
|
| |
|