| Model Number |
R SERIES |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including power cycling, bench handling and stress testing without duplicating the report.The color lcd display cable was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device's display turned completely white and no clinical information could be seen.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Search Alerts/Recalls
|