| Model Number |
E2516-GNSB |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during procedure, pencil caught on fire like a candle flame as resident was holding it right after using it on a patient.The flame was extinguished and removed from the field.Replaced it with new similar device and it worked fine which completed the procedure.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The pencil was received with an electrode.The investigation of the returned equipment did not identify anything that would have caused or contributed to the reported event.The investigation found no evidence of excessive heating or sparking on the returned pencil.The device was plugged into a generator and activated with satisfactory results.No abnormal sparking occurred.The device passed hipot testing indicating no electrical leakage.The investigation found the device to function normally and within specifications.The instructions for use (ifu) states, both oxygen and nitrous oxide support combustion.Watch for enriched oxygen and nitrous oxide atmospheres near the surgical site, especially during head and neck surgery.Enriched oxygen atmospheres may result in fires and burns to patients or surgical personnel.Do not exceed the maximum power limits as stated in the ifu.Exceeding these power limits may result in patient injury or product damage.If the electrode is damaged, discard it.Always use the lowest power setting to achieve the desired surgical effect.The maximum power settings are detailed in the instructions for use.Some surgeons may elect to buzz t he hemostats during a surgical procedure.This practice is not recommended and the hazards of such a practice probably cannot be eliminated.To minimize the risk, when using coated electrodes, place the edge of the electrode against the hemostat or metal instrument to assure a good electrical pathway.Do not activate the generator until the accessory makes contact with the instrument.Additional measures to minimize the risk of performing this technique are listed in the electrosurgical generator users manuals.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|