Catalog Number 1008197-20
Device Problems Difficult to Remove (1528); Stretched (1601); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/11/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device has not yet been received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat a non-tortuous, mildly calcified iliac artery.A 6.0x20mm viatrac balloon dilatation catheter failed to advance on a 0.014 guide wire (gw) as it met resistance with the gw and got stuck.The device was removed and it was noted that the tip was damaged.No patient effects.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Device codes: 2017 labeled.Internal file number - (b)(4).Corrections: device status changed from returning to not returned.Device code 2017-excessive force.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.It should be noted that the viatrac plus instruction for use states: if resistance is met during manipulation, determine the cause of the resistance before proceeding.The investigation was unable to determine a cause for the reported difficulties.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
Correction to case details: the physician applied excessive force to try to advance the balloon over the wire.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX VIATRAC 14 PLUS PERIPHERAL DILATATION CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8302281
MDR Text Key134910532
Report Number2024168-2019-00802
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
PMA/PMN Number
K102705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number1008197-20
Device Lot Number8011661
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/11/2019
Initial Date FDA Received02/04/2019
Supplement Dates Manufacturer Received02/28/2019
Supplement Dates FDA Received03/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age23 YR
-
-