| Model Number |
MMT-7020A |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose value was 117 mg/dl and the sensor glucose was 85 mg/dl.Insulin delivery was suspended due to sensor values.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.Customer states insulin pump was not working.Customer reports they did not receive a sensor updating alert.Customer reports they did not receive a calibration not accepted alert.The sensor will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened or used sensor and performed continuity resistance test and sensor passed per specifications.Performed the sensor initialization test using a new lab transmitter with a paradigm pump.Connected the sensor to the transmitter and placed it in 100 bts, confirmed the communication icon was displayed and that the transmitter was flashing while connected to the sensor the sensor functioned properly.Cannot performed bts test as the sensor is beyond its expiration date.
|
| |
|