ST. JUDE MEDICAL ADVISOR¿ FL, CIRCULAR MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
| Model Number |
D-AVSE-DF10-F20 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
| Date of Event |
01/14/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, cardiac perforation is an inherent risk of any electrode placement.
|
| |
|
Event or Problem Description
|
|
Related manufacturing ref: 3008452825-2019-00041, 3008452825-2019-00043, 3008452825-2019-00044, 3005334138-2019-00050, 3005334138-2019-00051, 3008452825-2019-00045, 3005334138-2019-00052.Following a pulmonary vein isolation procedure, a pericardial effusion was noted via echocardiography.A pericardiocentesis was performed to stabilize the patient.No patient symptoms were noted throughout the procedure and there were no performance issues with any abbott device.
|
| |
|