| Model Number |
8888411405 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led evaluation of one video of the device.The visual inspection video depicts the catheter being used during a surgical procedure.The surgeon is picking at the cuff.A review of the device history record could not be performed because the lot number was not provided.However, records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Our investigation was unable to determine a root cause or establish a relationship between the device and the reported condition.No enhancements or improvements were generated for the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, it was found that the cuff appeared to be out f the implant site.It was also found that the cuff slides (velvet an glue detached) and the glue was less compared to before.It was also stated that implant previously had an infection and there were concern there was damage to the tubing.There was no reported patient injury.
|
| |
|