| Model Number |
8785 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider (hcp) via a manufacturer representative on (b)(6) 2019 regarding a catheter intended for use with an implanted infusion pump.It was reported that an issue happened during surgery where three collets did not work properly and would not clamp down as much as the hcp would like.It was noted that they also split open on the other side instead of getting closed.One side did work properly, and the other side did not.No patient symptoms were reported and no further complications were reported or anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of catheter collet (b)(4) revealed the following: analysis determined the collet was not fully locked into place as received.The evaluation codes have been updated for catheter collet (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The catheter collet was returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Fdc updated to 19.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was indicated that all devices that were intended to be returned had been sent back, and the rest had been discarded.When asked about the cause of the issues with the collet not working properly, the representative stated the cause was unknown.
|
| |
|