MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ASCENDA; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ASCENDA; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8785
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/15/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) via a manufacturer representative on (b)(6) 2019 regarding a catheter intended for use with an implanted infusion pump.It was reported that an issue happened during surgery where three collets did not work properly and would not clamp down as much as the hcp would like.It was noted that they also split open on the other side instead of getting closed.One side did work properly, and the other side did not.No patient symptoms were reported and no further complications were reported or anticipated.
 
Additional Manufacturer Narrative
Evaluation of catheter collet (b)(4) revealed the following: analysis determined the collet was not fully locked into place as received.The evaluation codes have been updated for catheter collet (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The catheter collet was returned for analysis.
 
Additional Manufacturer Narrative
Fdc updated to 19.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was indicated that all devices that were intended to be returned had been sent back, and the rest had been discarded.When asked about the cause of the issues with the collet not working properly, the representative stated the cause was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCENDA
Common Device Name
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key8305333
Report Number3004209178-2019-02440
Device Sequence Number1619633
Product Code LKK
UDI-Device Identifier00763000051150
UDI-Public00763000051150
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P860004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model Number8785
Device Catalogue Number8785
Device Lot NumberHG2ZWE3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2019
Supplement Date Received by Manufacturer04/22/2019
04/22/2019
Initial Report FDA Received Date02/05/2019
Supplement Report FDA Received Date04/22/2019
04/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-