Catalog Number 71420188
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 01/14/2019
Type of Reportable Event Serious Injury
Event or Problem Description
A revision of a journey uni knee was performed and successfully revised it to a primary legion p/s knee.
 
Additional Manufacturer Narrative
"the associated complaint devices were returned and evaluated.The lab analysis concluded, the presence of bone cement left in the femur component indicates implant fixation in vivo.The scratches, dents, discoloration, and deformation observed do not present sufficient information to determine the root cause of the required revision surgery.No material or manufacturing deviations were observed in the process of this investigation.The clinical/ medical team concluded, although the asd product complaint form indicated: ¿x-rays to be sent with retrieval implants¿, no images have been provided to the medical investigation team as of the date of this assessment.Per the complaint details and communication received, the surgeon did not fault the device as it was reported that the root cause of the revision was due to an injured mcl status post a traumatic fall.The product evaluation reported ¿no material or manufacturing deviations¿ were observed and did not contribute further to the medical investigation.The patient impact beyond the revision procedure could not be determined.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaint for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical analysis indicated that this case reports that a revision surgery was performed to revise a journey uni knee to a primary legion p/s knee.Details regarding the patient¿s weight-bearing status, medical history, bone quality, fall/trauma history, and other additional clinical relevant information have not been provided.Based on the e-mails with smith and nephew sales rep, the surgeon does not fault the product, reportedly the patient fell and injured their mcl.The future impact to the patient cannot be determine.No further clinical assessment is warranted at this time.A review of complaint history on the listed parts revealed no prior complaints for the listed batches.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II CMT TIB SIZE 6 RIGHT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8305926
Report Number1020279-2019-00437
Device Sequence Number16552763
Product Code JWH
UDI-Device Identifier00885556029138
UDI-Public00885556029138
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71420188
Device Lot NumberJ1809827
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2019
Initial Date Received by Manufacturer 01/14/2019
Supplement Date Received by Manufacturer01/14/2019
11/25/2019
Initial Report FDA Received Date02/05/2019
Supplement Report FDA Received Date06/23/2019
11/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-