• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. COCR 12/14 FEM HD 26 + 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. COCR 12/14 FEM HD 26 + 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 71302600
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 01/10/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to recurrent dislocation.
 
Additional Manufacturer Narrative
The associated tandem bipolar shell and cobalt chrome femoral head were not returned for evaluation.Therefore a product analysis could not be performed.However, device details were provided.The review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.There is no information that would suggest the implanted devices failed to meet specifications.A relationship, if any, between the devices and the reported incident or adverse event could not be corroborated.Our clinical evaluation noted that no relevant supporting clinical information has been provided to assist with a clinical investigation.Therefore based on insufficient information, a thorough clinical assessment cannot be performed at this time.Without the return of the actual product involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Should the devices or additional information be received, the complaint will be reopened.
 
Event or Problem Description
It was reported that a revision surgery was performed due to recurrent dislocation.Primary surgery was performed on (b)(6) 2018 due to femoral neck fracture.After the primary surgery, posterior dislocation occurred twice on unknown dates, both were fixed by manual reduction.On (b)(6) 2019 a revision surgery was performed.The surgeon does not blame the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HD 26 + 0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8305992
Report Number1020279-2019-00401
Device Sequence Number71033
Product Code LPH
UDI-Device Identifier03596010194077
UDI-Public03596010194077
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K963509
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71302600
Device Lot Number18DM18419A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/11/2019
Supplement Date Received by Manufacturer01/11/2019
01/11/2019
Initial Report FDA Received Date02/05/2019
Supplement Report FDA Received Date04/09/2019
07/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-