MEDOS INTERNATIONAL SàRL CH VIPER TI SAI POLY 9X100MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 179704999
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the pelvic fixation surgery was performed on (b)(6).When the surgeon inserted the screw from the patient's left sacrum into the ilium, the surgeon recognized the strange noise, and then found the broken fragment (like screw thread) in the screw head.Thus, the surgeon removed the broken fragment.Then, the surgeon attempted final tightening with set screw; however, the set screw was spin idled.The surgery was completed by using alternative screw (p/n and lot # were unknown).There was less than 30 min.Surgical delay and there was no adverse consequence to the patient.No further information was provided by the hospital.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4).Visual examination found that the threads on the screw¿s tulip head was completely torn off.Noted damage suggests that inadvertently cross threading of the setscrew occurred upon insertion into the tulip head.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.With the information provided, a definitive root cause for the torn threads on the viper ti sai poly 9x100mm cannot be determined.Noted damage suggests that inadvertently cross threading of the setscrew occurred upon insertion into the tulip head.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER TI SAI POLY 9X100MM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key8307023
Report Number1526439-2019-51312
Device Sequence Number1619717
Product Code NKB
UDI-Device Identifier10705034265889
UDI-Public(01)10705034265889
Combination Product (Y/N)N
PMA/510(K) Number
K111571
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number179704999
Device Catalogue Number179704999
Device Lot NumberTBTWA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2019
Initial Date Received by Manufacturer 01/16/2019
Supplement Date Received by Manufacturer04/02/2019
Initial Report FDA Received Date02/05/2019
Supplement Report FDA Received Date04/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-