• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP¿ FOLEY TRAY SYSTEM WITH BARD® LUBRI-SIL® TEMP SENSING FOLEY CATHETER; SURESTEP FOLEY CATHETER TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP¿ FOLEY TRAY SYSTEM WITH BARD® LUBRI-SIL® TEMP SENSING FOLEY CATHETER; SURESTEP FOLEY CATHETER TRAY Back to Search Results
Model Number A119216M
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the surestep catheter tray was missing the iodine swabsticks, the lube syringe and the sterile water syringe.The remaining items of the tray were used with no patient impact.No medical intervention was reported.
 
Event or Problem Description
It was reported that the surestep catheter tray was missing the iodine swabsticks, the lube syringe and the sterile water syringe.The remaining items of the tray were used with no patient impact.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was confirmed.Visual evaluation of the returned sample noted one opened tray in which no syringes or iodine swabs were present.One foam swab was present.It was noted that the product should contain iodine packets and foam swabs rather than iodine swabs.Missing components do not comply with the standard and is out of specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿proper techniques for urinary catheter insertion ¿ perform hand hygiene immediately before and after insertion ¿ insert urinary catheters using aseptic technique and sterile equipment ¿ use the smallest foley catheter possible, consistent with good drainage ¿ document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient record proper techniques for urinary catheter maintenance ¿ secure the foley catheter, use the statlock® foley stabilization device if provided ¿ maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions ¿ maintain unobstructed urine flow and keep the catheter and collection tube free from kinking ¿ keep the collection bag below the level of the bladder or hips at all times ¿ empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient ¿ routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate ¿ leave foley catheter in place only as long as needed".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESTEP¿ FOLEY TRAY SYSTEM WITH BARD® LUBRI-SIL® TEMP SENSING FOLEY CATHETER
Common Device Name
SURESTEP FOLEY CATHETER TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8307076
Report Number1018233-2019-00605
Device Sequence Number62206
Product Code EZL
UDI-Device Identifier00801741073687
UDI-Public(01)00801741073687
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2021
Device Model NumberA119216M
Device Catalogue NumberA119216M
Device Lot NumberNGCX0604
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2019
Initial Date Received by Manufacturer 01/14/2019
Supplement Date Received by Manufacturer04/24/2019
Initial Report FDA Received Date02/05/2019
Supplement Report FDA Received Date05/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-