• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ORTHOMATCH MOD FEM TRIAL IMPACTOR; PROSTHESIS KNEE PATELLOFEMOROTIBIAL SEMI-CONSTRAINED CEMENTED POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ORTHOMATCH MOD FEM TRIAL IMPACTOR; PROSTHESIS KNEE PATELLOFEMOROTIBIAL SEMI-CONSTRAINED CEMENTED POLY/METAL/POLY Back to Search Results
Catalog Number OR71442161
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/14/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during procedure the impactor broke.All pieces accounted for.No delay or injury reported.Procedure was completed with the same device.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The modular connector fractured off of the device and was returned.The impactor body was heavily damaged from repeated impacts.The device was manufactured in 2016 and exhibits signs of extensive wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORTHOMATCH MOD FEM TRIAL IMPACTOR
Common Device Name
PROSTHESIS KNEE PATELLOFEMOROTIBIAL SEMI-CONSTRAINED CEMENTED POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8307635
Report Number1020279-2019-00415
Device Sequence Number13468227
Product Code JWH
UDI-Device Identifier00885556583791
UDI-Public00885556583791
Combination Product (Y/N)N
PMA/510(K) Number
K142807
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberOR71442161
Device Lot Number16FBG0033
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 01/15/2019
Supplement Date Received by Manufacturer01/15/2019
Initial Report FDA Received Date02/05/2019
Supplement Report FDA Received Date03/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-