| Catalog Number |
011020 |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
01/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy the pinch broke inside the patient's knee.Then they tried several times to remove the piece but did not succeed.A check was done by x-rays (intraoperative).This has lead to the opening of the knee, an increase of the withers time, 15 minutes of operative arthoscopic gesture for 1h10 minutes of opening time and antibioprophylaxy.It is unknown if a back up device was available.A delay greater than 2 hours was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One 2.7mm alligator grasper was returned for evaluation.The device is 23 yrs.Old and exhibits expected wear and tear.Visual assessment of the device confirmed the reported complaint.The upper jaw has returned and has broken off where it transitions into the lug.The lug portion of the jaw remains attached to the shaft.The normally sharp teeth of the jaws are rounded over.The condition of the device indicates it was subjected to excessive force during use.Per the devices instructions for use under precautions: ¿as with any surgical instrument, careful attention should be exercised to ensure that excessive force is not placed on the instrument.Excessive force can result in instrument failure.¿.
|
| |
|