BARD ACCESS SYSTEMS POWERPORT MRI ISP W/O SUTURE PLUGS, 9.6 FR SILICONE SL CATHETER; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 1859661
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 9/2018).The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that approximately eight months after port-a-catheter placement during a chest x-ray; the x-ray demonstrated the left subclavian port-a-catheter was fractured with the distal half of the catheter over the right cardiac border.It was further reported the port and the proximal catheter in left chest was removed without complications.Reportedly, the distal segment was retained in the right atrium and was removed a few days after the port was removed.The patient status is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT MRI ISP W/O SUTURE PLUGS, 9.6 FR SILICONE SL CATHETER
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8308451
Report Number3006260740-2019-00175
Device Sequence Number53502
Product Code LJT
UDI-Device Identifier00801741027239
UDI-Public(01)00801741027239
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K073423
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1859661
Device Catalogue Number1859661
Device Lot NumberREBT0508
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2019
Initial Report FDA Received Date02/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient Weight93
-
-