SMITH & NEPHEW, INC. TF ULTRA 2 UBRAID AND NDLS 5.5MM PK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72202615 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a shoulder rotator cuff repair surgery, the non-bioabsorbable twinfix ultra suture anchor tip broke while inserted into the patient bone.A back-up was available and used in the same bone hole, but faced the same faith.All fragments were removed from the patient.There was a surgical delay of less than 30 min.No patient injuries have been stated.For preparation of the insertion site, it was used a 3.8 mm gold dilator; patient bone quality was described as hard.
|
| |
|
Additional Manufacturer Narrative
|
|
One (b)(4) twinfix ultra pk 5.5mm suture anchor reported on.Product was not returned for evaluation.Due to unavailability, the allegation could not be fully confirmed.Full investigation and conclusions were not possible without physical evaluation.If objective evidence becomes available to assist with evaluation, the complaint will be revisited.
|
| |
|