Catalog Number EVD35-06-120-150
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 01/25/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Physician used an everflex entrust self-expanding stent with a 6fr non-medtronic (cook ansel) sheath and 0.035 wholey wire during treatment of a calcified lesion in the patients proximal and mid left external iliac to left common femoral artery.Slight tortuous and moderate calcification is reported.Ifu was followed and the device was prepped without issue.The lock-pin was checked for securement prior to the procedure.Pre-dilation was performed using a 5x40 pre-dilation device.The device was not passed through a previously deployed stent.No resistance was encountered during advancement.The lock-pin was removed prior to deployment.It is reported following deployment of approx.2cm of the stent it jumped dramatically.The 120mm stent appeared to be only 80mm when deployed.The stent was post dilated and a second stent was deployed without difficulty.No patient injury reported.
 
Additional Manufacturer Narrative
Device evaluation: the everflex entrust stent delivery system, (sds), was received for evaluation within a sealed tyvek pouch, within a sealed biohazard pouch, and loosely coiled within a red plastic biohazard pouch tied shut.No ancillary devices from the procedure were received for evaluation.The entrust was received with the outer sheath retracted and the inner guidewire lumen/pusher exposed.The deployment handle was examined: the silver outer sheath is visible over the inner guidewire lumen in the safety tab cavity.Approximately 12.6cm of the inner guidewire lumen is exposed.The printed strain relief was removed from the deployment handle to allow further examination of the handle.The handle was split opened, the pull cable was fully wrapped around the thumbwheel, and the silver outer sheath proximal end was advanced proximal to the pulley.The pull cable to outer sheath bond was broken.The pull cable to outer sheath tensile strength bond is equal to or greater than 8.0lbs.The pull cable most likely separated from the outer sheath after stent deployment and after the outer sheath was fully retracted into the handle based on the damage to the outer sheath.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information received reported that approximately 2cm of the stent jumped forward.Stent jumped in the targeted area.No patient injury or vessel damage.Second stent was placed due to the stent jumping.No other intervention was required.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
MDR Report Key8309591
Report Number2183870-2019-00057
Device Sequence Number19425870
Product Code NIP
UDI-Device Identifier00821684051627
UDI-Public00821684051627
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
P110023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2021
Device Catalogue NumberEVD35-06-120-150
Device Lot NumberA748302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2019
Initial Date Received by Manufacturer 01/25/2019
Supplement Date Received by Manufacturer02/06/2019
04/19/2019
Initial Report FDA Received Date02/06/2019
Supplement Report FDA Received Date03/04/2019
04/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-