• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS INSULIN CARTRIDGE; INSULIN PUMP CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS INSULIN CARTRIDGE; INSULIN PUMP CARTRIDGE Back to Search Results
Lot Number D200494
Medical Device Problem Code Failure to Prime (1492)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.Each cartridge lot is subjected to a statistical sampling plan and must pass testing for force (occlusion and loss of prime), cracks, and foreign material prior to release for distribution.
 
Event or Problem Description
On (b)(6) 2019, the reporter contacted animas, alleging a prime (loss of prime) issue.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS INSULIN CARTRIDGE
Common Device Name
INSULIN PUMP CARTRIDGE
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key8310143
Report Number2531779-2019-00924
Device Sequence Number16736913
Product Code LZG
UDI-Device Identifier10840406100129
UDI-Public10840406100129
Combination Product (Y/N)Y
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K032257
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Lot NumberD200494
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2019
Initial Report FDA Received Date02/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age15 YR
-
-