|
The concerto coil was not returned for analysis; therefore, no definitive conclusion can be drawn regarding the clinical observation.Per the concerto¿ detachable coil and i.D.(instant detacher) instructions for use (ifu): if the coil does not detach after 3 attempts, discard the i.D.(instant detacher) and replace with a new i.D.(instant detacher).Also, ¿in the rare event that the coil does not detach and cannot be removed from the implant delivery pusher, use the following steps for detachment.Grip the hypotube approximately 5 cm distal of the positive load indicator at the hypotube break indicator and bend the implant delivery pusher just distal to the hbi 180 degrees.Next straighten the pusher back, continue bending and straightening until the pusher tubing opens exposing the release element.Gently separate the proximal and distal ends of the open pusher.Then, under fluoroscopy, pull the proximal portion of the implant delivery pusher approximately 2-3 cm to confirm implant detachment per ifu.¿ if information is provided in the future, a supplemental report will be issued.
|
|
The pusher was returned only; within its inner pouch; inside of a biohazard bag and a shipping box.There was no implant coil, instant detacher, microcatheter, or accessories returned with the device.Kinks and bends are present along the length of the pusher ~36.0 to 139.0cm from the proximal end.The ai to release wire crimps, coupler to ai crimps, and coupler to pusher tube weld are present and loose; it appeared that mechanical detachment was attempted.The pusher was found broken at the break indicator; but retained by the release wire; it appeared that the manual detachment was attempted at this location.The coin was not present at the lumen stop as it was pulled back.The shield coil was present and not stretched.Under the microscope, the outer jacket was then removed to gain access the coin.The coin was measured in 3 locations and was found to be within specifications.Measured 0.075mm @ 0.063mm; measured 0.084mm @ 0.127mm; measured 0.092mm @ 0.275mm.The inner diameter of the lumen stop and the inner diameter of the retainer ring were found to be visually acceptable.The lumen stop inner diameter (id) was measured to be 0.00260¿ and found to be within specification (0.00220" minimum - 0.0029" maximum).The retainer ring inner diameter (id) was measured to be 0.00455¿and found to be within specification (0.00455"+/- 0.00010").All other subassemblies appeared to be normal.No other anomalies were observed.Based on the analysis performed the concerto coil was confirmed to have prematurely detached.There was no malfunction of the pusher or non-conformance to specification that may have contributed to the pre-mature detachment.Since the implant coil, detached element, instant detacher and microcatheter were not returned, any contribution of the implant coil, detached element, instant detacher or microcatheter to the issue could not be determined.Based on the event description it is was not possible to ascertain the cause of the pre-mature detachment; however, based on the returned device, there was evidence of mechanical and manual detachment attempts to detach the coil.The pusher was also found to be broken and retained by the release wire; with the coin pulled back from the lumen stop.The pulling back of the coin from the lumen stop m ay have contributed to the detachment of the implant coil from the push wire.Furthermore, the pusher was bent and kinked along its length, indicative of high tortuosity.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|