Model Number DM3500
Device Problem Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the mobile application had missing device alert checks since (b)(6) 2019.Upon review, the insertable cardiac monitor reset had occurred and that monitoring was disabled.The patient would follow up for the device interrogation.No patient symptoms were reported.
 
Manufacturer Narrative
Device was returned.Analysis of the device image indicated that memory corruption occurred and triggered the alert message.Further testing was performed by reloading the product code and all tests results were normal, memory corruption could not be reproduced.The reset error noted is consistent with that occurred due to discrepant ic.
 
Event Description
New information received noted that the patient presented for a device interrogation.Upon review, the device was unable to be interrogated.The device was explanted and replaced on (b)(6) 2019.The patient was stable after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8310700
MDR Text Key135177707
Report Number2017865-2019-01804
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/25/2019
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000059076
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
-
-