| Model Number |
R11278AUK1 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
We evaluated the subject flexible scope and we found multiple cuts and tooling/grab marks throughout the length of the angle cover.The damage begins at the distal thread wrap area and continues back to the proximal thread wrap.It appears this is damage caused by instruments used during the procedure.The distal head weld had completely separated on one side and partially separated on the other.This type of damage would take excessive force.The flexible ureteroscope has been in use for approximately 5 months.
|
| |
|
Event or Problem Description
|
|
During a ureteroscopy procedure the distal wrap of the scope came apart inside the patient's bladder.The broken piece was retrieved and the procedure was completed with no impact on the patient.
|
| |
|