| Model Number |
MT22719-PNK |
| Medical Device Problem Code |
Inappropriate or Unexpected Reset (2959)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2019 that on (b)(6) 2019 the receiver initialized without a manual restart.No additional event or patient information is available.No product or data was provided for evaluation.The confirmation of the complaint is undetermined.The probable cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Com-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver initialized without manual restart.The receiver was returned for evaluation.An external visual inspection was performed and passed.A receiver charge and boot was performed and failed due to perpetual booting.A functional test was not performed.The log was downloaded and reviewed finding errors related to the complaint in receiver shutdown at high battery capacity (level 60) followed by perpetually rebooting.The receiver case was opened, and an internal visual inspection was performed and failed due to damaged battery.Confirmation of the complaint was confirmed.No injury or medical intervention was reported.
|
| |
|