• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 030450
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic video-assisted thoracoscopic lobectomy surgery, while on the inferior pulmonary vein, three loading units broke off.The operation technique was changed and converted to open procedure that the operating time was delayed for 30 minutes or more.Extended incision was also reported and another loading unit was used to complete the case.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of a photo and one device.The proximal ends of the adapters were broken.A review of the device history record indicates the product was released meeting all quality release specifications.Replication of the damaged loading units adapters may occur due to over flexure of the loading units while attached to the instrument.The root cause of the observed damages were misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8315577
Report Number1219930-2019-00750
Device Sequence Number16410086
Product Code GDW
UDI-Device Identifier10884523002812
UDI-Public10884523002812
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number030450
Device Catalogue Number030450
Device Lot NumberT8D007X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2019
Initial Date Received by Manufacturer 01/15/2019
Supplement Date Received by Manufacturer04/23/2019
Initial Report FDA Received Date02/07/2019
Supplement Report FDA Received Date05/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age32 YR
Patient Weight90
-
-