MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910  DURALOC SECTOR SERIES 50OD; DURALOC IMPLANT : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910  DURALOC SECTOR SERIES 50OD; DURALOC IMPLANT : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 124580056
Device Problem Defective Device (2588)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/14/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Size issue.The outer diameter of acetabular cup is too small to implant.There is a little bit smaller than label indicates as the photo shows.Another device (52mm) was used to complete the surgery.There were no adverse consequences to the patient.
 
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: the returned part was examined by the manufacturer and the device met specification.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURALOC SECTOR SERIES 50OD
Type of Device
DURALOC IMPLANT : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8315808
MDR Text Key135383633
Report Number1818910-2019-83594
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295019657
UDI-Public10603295019657
Combination Product (y/n)N
PMA/PMN Number
K961186
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial,Followup,Followup
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number124580056
Device Lot NumberD18033044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2019
Date Manufacturer Received04/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-