| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 10-oct-2018 with the following findings: during investigation, the battery compartment was found to be cracked.(b)(6).
|
| |
|
Event or Problem Description
|
|
The pump was returned for investigation.Investigation revealed a battery compartment crack.This report is made based on results of investigation completed on 10-oct-2018.
|
| |
|
Additional Manufacturer Narrative
|
|
The date that product analysis completed their investigation was on 02-oct-2018.
|
| |
|
Search Alerts/Recalls
|