| Catalog Number |
K1039 SMN 10445162 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens healthcare diagnostics headquarters support center (hsc) concluded their investigation of the discordant, falsely depressed glucose result.The customer noted that the sample had a black discoloration.This was the only sample with this issue.The customer processed other samples without issue.Quality control (qc) was reprocessed after this patient sample and it was within laboratory ranges.Hsc concluded that this is not a reagent lot or assay issue.There is no evidence of a product nonconformance.There was no indication of foaming or reagent delivery issues based on the instrument data files.A siemens customer service engineer (cse) was dispatched to the customer account.Product maintenance was performed and no further discordant glucose results were observed.The device is performing within specifications.No further evaluation is required.
|
| |
|
Event or Problem Description
|
|
A discordant, falsely depressed glucose (glu) result was obtained on a patient sample on the dimension vista 500 system.The discordant result was not reported to the physician.The same sample was reprocessed the same day with an alternate, non-siemens methodology and a higher result was obtained and reported.There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely depressed glucose result.
|
| |
|