• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. ALTIVATE SHOULDER; AR, SMALL SOCKET INSERT, 40MM NEUTRAL +4 EPLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. ALTIVATE SHOULDER; AR, SMALL SOCKET INSERT, 40MM NEUTRAL +4 EPLUS Back to Search Results
Model Number 509-02-440
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/21/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to instability and dislocation.The surgeon need to increase poly for instability.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to instability.The previous surgery and the surgery detailed in this investigation occurred 21 days apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to instability.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE SHOULDER
Common Device Name
AR, SMALL SOCKET INSERT, 40MM NEUTRAL +4 EPLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8317314
Report Number1644408-2019-00094
Device Sequence Number59420
Product Code KWS
UDI-Device Identifier00190446257653
UDI-Public(01)00190446257653
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K172351
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/08/2022
Device Model Number509-02-440
Device Catalogue Number509-02-440
Device Lot Number961W1002
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/21/2019
Supplement Date Received by Manufacturer02/13/2019
Initial Report FDA Received Date02/07/2019
Supplement Report FDA Received Date02/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
-
-