| Model Number |
MMT-551PNAL |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via call that the top of the reservoir compartment was damaged.Customer¿s blood glucose level was unknown at the time of incident.Customer state that no longer holds the reservoir.The insulin pump will be returned for the analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform displacement test due to completely broken reservoir tube lip.Device had missing reservoir tube o-ring.
|
| |
|