• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is (b)(6) 2018.(b)(4).All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens was explanted from patient¿s left eye in secondary surgical procedure because of patient experiencing poor visual acuity.Reportedly there was no incision enlargement, no vitrectomy, no sutures required.Patient outcome is good.No further information provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 2/1/2019.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The product was received dry in a little jar.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint can be not confirmed.Manufacturing record review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specifications.The search revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8319245
Report Number9614546-2019-00105
Device Sequence Number19211561
Product Code POE
UDI-Device Identifier05050474579200
UDI-Public(01)05050474579200(17)211112
Combination Product (Y/N)Y
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/09/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/12/2021
Device Model NumberZXR00
Device Catalogue NumberZXR00U0255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2019
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 01/15/2019
Supplement Date Received by Manufacturer02/11/2019
10/25/2020
Initial Report FDA Received Date02/08/2019
Supplement Report FDA Received Date02/15/2019
11/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-