| Catalog Number |
0620040600 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was over inflating with no alarms.Please note, the procedure was completed successfully.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was over inflating with no alarms.Please note, the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cib jennifer surgery over inflating and no alarm, po# (b)(4) the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.There is no probable root cause in relation to the complaint as the unit performed to specification with no errors observed as per the customer complaint.However, the unit is past due for calibration and preventative maintenance.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|