| Model Number |
BL5223W |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sterilization and lot history records were reviewed and found to be acceptable.Further investigation underway.Report 2 of 2.See 1920664-2019-00009.
|
| |
|
Event or Problem Description
|
|
The user facility in (b)(6) reported the vitrectomy cutter could not cut during surgery.Aspiration continued when the cutter stopped.There was no impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: device manufacture date.The product evaluation is completed: one opened bl5223w pack from lot v9914 was returned.The expiration date on the label was 2019-jan-22.The tubing was still coiled and taped together.The assembly was clean and dry.This pack does not appear to have been used.The tip protector was in place, as it should be.The needle was not bent.The port window was in the opened position.There was no evidence of dried solution visible in the tubing.The back cap was aligned correctly with the vent hole in the side of the cutter body.The connectors were inspected and no defects were found.A functional test was performed using a stellaris pc system.The cutter had good clean cuts throughout the various cut rates and performed as intended.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Capa 610815 was closed on 06 july 2018 but was still open when lot v9914 was manufactured.Per capa 610815 the vitrectomy tubeset used on the stellaris pc packs and standalone pouched vitrectomy cutters was replaced with a new vitrectomy tubeset.The new vitrectomy tubeset provides a greater operating range between the pressure required to close the vitrectomy cutter inner needle and the pressure required to open the vitrectomy cutter inner needle.The vitrectomy tubeset enhancement was implemented on the impacted pc packs and pouched accessories, including bl5223w.The first lot of bl5223w that was built with the vitrectomy tubeset enhancement was lot #w2087 in (b)(6)2018.Any bl5223w product with lot no.W2087 or higher will include the vitrectomy tubeset enhancement.A review of the complaint database revealed no other complaints have been submitted against lot# v9914.Investigation complete.Report 2 of 2 see 1920664-2019-00009.
|
| |
|
Search Alerts/Recalls
|