Catalog Number 306572 |
Device Problem
Insufficient Information (3190)
|
Patient Problems
Dyspnea (1816); Swelling (2091)
|
Event Date 12/14/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 8179538, medical device expiration date: 2021-06-30, device manufacture date: 2018-06-28.Medical device lot #: 8185572, medical device expiration date: 2021-06-30, device manufacture date: 2018-07-04.Medical device lot #: 8242797, medical device expiration date: 2021-08-31, device manufacture date: 2018-08-30.(b)(6).
|
|
Event Description
|
It was reported with the use of the bd posiflush¿ xs pre-filled flush syringe there was an issue with patient appearing breathless with swelling to both legs.
|
|
Event Description
|
It was reported with the use of the bd posiflush¿ xs pre-filled flush syringe there was an issue with patient appearing breathless with swelling to both legs.
|
|
Manufacturer Narrative
|
Investigation summary: dhr: the non-conformances were reviewed for this batch, and there was no record of non-conformance which could contribute to the complaint verbatim reported by the customer.No sample available.Conclusion(s): there is no evidence that posiflush syringe was responsible for this reaction.A potential contributory factor maybe other medicinal products in use or administered at the time the patient¿s port was flushed.Conclusion(s): based on the information provided, it is more probable than not that the symptoms described may be an allergic reaction; however, it is highly improbable that this reaction was produced by the normal saline in the bd posiflush product.During the period 2017-2019, there are no other adverse customer complaint trends for the complaint category of allergic reaction, apart from (b)(4) uk.There are no known allergies to normal saline, either topical or within the body.It is most likely that whatever fluid was in the iv line prior to flushing contained inadequately flushed medication or glucose from previous treatments.
|
|
Search Alerts/Recalls
|
|
|