Catalog Number 72202625
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/30/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery, the anchor broke while screwed in.All the pieces were removed.A backup device was available to complete the procedure with no significant delay and no patient injuries.
 
Additional Manufacturer Narrative
Additional information.Correction.
 
Event or Problem Description
It was reported that during a shoulder surgery, the anchor broke while screwed in.All the pieces were removed.There was no loss of tissue or bone due to the event.A backup device was available to complete the procedure with no significant delay and no patient injuries.
 
Additional Manufacturer Narrative
One 5.5 twinfix ultra pk anchor assembly was returned for evaluation.Visual assessment of the anchor confirmed the reported breakage.The anchor has broken at the suture eyelet.Dimensional assessment of the anchors major diameter found it met print specifications.The inserter tines that interface with the anchor were also examined and noted to be intact.The condition of the anchor indicates it was subjected to excessive force during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA PK 5.5MM W/ 2 UB-BL &BLK
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8325168
Report Number1219602-2019-00170
Device Sequence Number19261750
Product Code MBI
UDI-Device Identifier03596010647535
UDI-Public03596010647535
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K093228
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72202625
Device Lot Number2013881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2019
Initial Date Received by Manufacturer 01/31/2019
Supplement Date Received by Manufacturer02/21/2019
03/11/2019
Initial Report FDA Received Date02/11/2019
Supplement Report FDA Received Date02/21/2019
03/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
-
-