Catalog Number 960081
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 01/23/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.(b)(4).
 
Event or Problem Description
It was reported by the surgeon that the patient was having anterior knee pain the could have been caused by the femur being internally rotated and subsequent poor patella tracking.The surgeon removed the femoral component and replaced with a new tc3 femur, correcting the rotation and moving the joint line more distill with thicker distal augments.A new insert was implanted.Doi: (b)(6) 2017; dor: (b)(6) 2019; left knee.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary : per internal procedures, the event information and any investigational inputs received as part of required follow up were reviewed.For this investigation, no immediate action was required as no alleged deficiency with the device was identified.Monitor exempt per (b)(4) known possible complication of joint replacement surgery no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.The reported event is considered one of the possible complications of joint replacement.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.Without the physical complaint sample associated with this report, it was not possible to determine if the device failed to meet specification at the time it was released for distribution.The device associated with this event was used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No product contribution to the reported event was identified.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMA TC3 FEM LEFT SZ2
Common Device Name
SIGMA REVISION IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8326644
Report Number1818910-2019-84036
Device Sequence Number13888159
Product Code JWH
UDI-Device Identifier10603295232483
UDI-Public10603295232483
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K952830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number960081
Device Lot Number616053
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/23/2019
Supplement Date Received by Manufacturer03/15/2019
Initial Report FDA Received Date02/11/2019
Supplement Report FDA Received Date03/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-