• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STRAIGHT SUCTION 9733449 EM ENT; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STRAIGHT SUCTION 9733449 EM ENT; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9733449
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A medtronic representative went to the site to test the system the straight suction was used with.Testing revealed that the system performed as intended and no hardware parts were replaced.The system then passed the system checkout and was found to be fully functional.The instrument has not been returned, so no analysis has been conducted on it.Refer to mdr 1723170-2019-00614 for the report on the first straight suction.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
(b)(6) 2019, (b)(4) - (hcp, rep) medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess) procedure.The issue occurred intraoperatively during the navigate task and caused no delay to the surgery.It was reported that the surgeon felt inaccurate with the straight suction (3 mm) but no other instrument.The surgeon aborted the use of the navigated suction and used a probe to confirm position.The medtronic representative noted that the suction did need slight manipulation to verify but only displayed a distance to divot around 2.1 mm before verification.The surgeon felt accurate after registration with an error metric of 2.1.The surgery was completed with navigation and there was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT SUCTION 9733449 EM ENT
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key8326974
Report Number1723170-2019-00615
Device Sequence Number2137928
Product Code HAW
UDI-Device Identifier00613994378422
UDI-Public00613994378422
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9733449
Device Catalogue Number9733449
Device Lot Number17060848
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2019
Initial Report FDA Received Date02/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
Patient Weight84
-
-