• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED SENSOR MMT-7020A ENLITE3 5PK US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED SENSOR MMT-7020A ENLITE3 5PK US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-7020A
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Calibration Problem (2890)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via that the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 82mg/dl and the sensor glucose was 50mg/dl insulin delivery was suspended due to sensor values.Sensor value that triggered the suspend event was 52mg/dl.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will be returned for analysis.
 
Additional Manufacturer Narrative
Inspected one opened and used sensor and performed visual inspection and found contact pad to be damaged (wrinkled).Unable to confirm customer received sensor in said condition due to customer returned opened and used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR MMT-7020A ENLITE3 5PK US
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8329292
Report Number2032227-2019-01361
Device Sequence Number13987777
Product Code OZP
UDI-Device Identifier20643169704057
UDI-Public(01)20643169704057(17)190409(10)D098
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/09/2019
Device Model NumberMMT-7020A
Device Catalogue NumberMMT-7020A
Device Lot NumberD098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2019
Initial Date Received by Manufacturer 07/10/2018
Supplement Date Received by Manufacturer03/13/2019
Initial Report FDA Received Date02/11/2019
Supplement Report FDA Received Date03/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
Patient Weight185
-
-