| Model Number |
MMT-523NAP |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor position encoder error.The customer¿s blood glucose was 287 mg/dl at the time of incident.Customer reports the drive support cap appears normal.The customer was advice the insulin pump will need to be replaced.Advice to discontinue use of the insulin pump and revert to backup plan per health care professional instructions.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with motion sensor test failure alarmed during rewind due to motor encoder signal out of phase.Unable to perform operating currents, self-test, unexpected restart error test, rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test or verify motor position encoder and motor error alarm due to motion sensor test failure alarms.
|
| |
|