Model Number MMT-523NAP
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had motor position encoder error.The customer¿s blood glucose was 287 mg/dl at the time of incident.Customer reports the drive support cap appears normal.The customer was advice the insulin pump will need to be replaced.Advice to discontinue use of the insulin pump and revert to backup plan per health care professional instructions.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with motion sensor test failure alarmed during rewind due to motor encoder signal out of phase.Unable to perform operating currents, self-test, unexpected restart error test, rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test or verify motor position encoder and motor error alarm due to motion sensor test failure alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8330476
Report Number3004209178-2019-54616
Device Sequence Number11088764
Product Code OYC
UDI-Device Identifier00613994743879
UDI-Public(01)00613994743879
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAP
Device Catalogue NumberMMT-523NAP
Device Lot NumberA2523NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2019
Initial Date Received by Manufacturer 02/09/2019
Supplement Date Received by Manufacturer04/23/2019
Initial Report FDA Received Date02/12/2019
Supplement Report FDA Received Date04/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age57 YR
Patient Weight162
-
-