| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.This report is made under the requirements of the medical device reporting regulations and does not constitute an admission on the part of animas of any deficiency in the performance of the device.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue; a cracked cartridge compartment.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction can result in a delay in treatment or long-term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
| |
|
Additional Manufacturer Narrative
|
|
Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Device analysis: the device was returned and investigation by product analysis completed on 01-may-2019 with the following findings: on investigation, the cartridge compartment was confirmed to be cracked.Investigation duplicated the complaint.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is made under the requirements of the health authority regulations and does not constitute an admission on the part of animas of any deficiency in the performance of the device.
|
| |
|