Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.The pump has been returned and evaluated by product analysis on 06-feb-2019 with the following findings: device evaluation: on investigation, there was contamination under the contacts of all the keypad buttons; however, all buttons were normally responsive.The battery compartment was also noted to be cracked.
 
Event or Problem Description
The pump was returned for investigation.Investigation revealed a keypad button issue and a case/condition issue.This report is made based on results of investigation completed on 06-feb-2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key8331044
Report Number2531779-2019-01177
Device Sequence Number16158717
Product Code OYC
UDI-Device Identifier10840406100068
UDI-Public10840406100068
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2019
Was the Report Sent to FDA? Yes
Device Age47 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2019
Initial Report FDA Received Date02/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age19 YR
-
-