| Model Number |
33000 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer did not provide patient demographics such as weight, ethnicity or race.The access b12 reagent was not returned for evaluation.All assay and system verifications met specifications at the time of this incident.No hardware errors, flags or other assay issues were reported in conjunction with this event.The cause of this event cannot be determined with the available information.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019 the customer reported that non-reproducible elevated vitamin b12 (access vitamin b12) results had been generated on the customer's unicel dxi800 immunoassay system (serial number (b)(4)) for one patient sample.The initial elevated vitamin b12 result was 219.23 pmol/l was released from the laboratory.The sample was sent to another laboratory and repeated with an alternate method (not provided); result provided was <111 pmol/l.There was no report of change to patient care or treatment.Calibration, quality control and system check were all performing within specifications at the time of the incident.Information regarding patient sample collection and processing was not provided.
|
| |
|