| Catalog Number |
68-150 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side "spontaneous ruptured implant one week." device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: opening, crease fold and wear abrasion.Leak test was performed and found opening on radius.A microscopic analysis was performed which identify a striated edge opening in the posterior side, consist with use of a surgical tool.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a striated edge opening on posterior side due to surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side "spontaneous ruptured implant one week." device has been explanted.
|
| |
|