• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENCORE 26; SYRINGE, BALLOON INFLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENCORE 26; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number M00566670
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation results: visual examination of the complaint device found that there was a white stain on the side of the device, inside the clear housing and on the outside of the barrel.The stain did not obstruct the graduation marks in the barrel.The stain is consistent with residues of dried adhesive used during the automated manufacturing process.The most probable root cause is manufacturing deficiency, as the design or validation of a suppliers manufacturing process was not sufficient to ensure the finished device met the intent of the design.An investigation to address this issue has been completed.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event or Problem Description
It was reported to boston scientific corporation that an encore inflation device was used in the bile duct during a biliary dilation procedure performed on (b)(6) 2019.According to the complainant, during preparation, a dirt was noticed in the inner tube of the inflator.The procedure was completed with another encore inflation device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE 26
Common Device Name
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8332381
Report Number3005099803-2019-00512
Device Sequence Number62905
Product Code MAV
UDI-Device Identifier08714729755241
UDI-Public08714729755241
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K955869
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/17/2021
Device Model NumberM00566670
Device Catalogue Number6667
Device Lot Number0022669104
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2019
Initial Report FDA Received Date02/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-